CR Pro Phosphorus Plate Calibration Procedure
Figure 3-1 – Calibrate Settings Screen
Expose Plate – 3 phase and high-frequency X-ray machines.
SID 72” - kVp 70 - mAs - 30
40" can also be used if 72" is impractical (e.g., because of the wall
bucky, etc). - 3 phase and high-frequency X-ray machines, reduce the 30
mAs exposure by a factor of approx 3, which is approx 10 mAs.