Installation guide

5
The
AC outlet should be dedicated to the CR-Pro (no other devices
connected to this circuit) to prevent the possibility of injected line “noise “by
other devices.
FUSE –
The input power is fused at the AC power input connector at the
bottom, right rear of the CR-Pro cabinet. The fuses are located in a box
directly above the input connector and are an integral part of the connector
assembly.
Inside the box are two (2) fuses as follows:
Domestic 100 – 120VAC 5A Fuses (standard)
European 200 – 240VAC 3A Fuses (standard)
ENVIRONMENT –
Temperature – 60ºF (19ºC) to 80ºF (32ºC) Operating
40ºF (4ºC) to 90ºF (33ºC) Non-operating
Relative Humidity – 20% to 80% non-condensing
Light – Phosphor plates are light sensitive and are erased if exposed to light.
The CR-Pro cassettes are light protected. However, light can penetrate the
CR-Pro light seal if bright light or sunlight is directed to the cassette slot on
top of the CR-Pro. Ambient light in the proximity of the CR-Pro should be kept
to a minimum.
SHIPPING AND UNPACKING THE CR-PRO –
The CR-Pro is
shipped in a custom designed box that protects the unit from shipping
damage. When the unit is returned for any reason, it must be package in the
original container. Retain all Interior protective parts and hardware for reuse.
NOTE –
Unit shall be unpacked and installed by authorized service
provider.
CAUTION –
The packing crate can be tipped when shipping and
moving and should be tethered to avoid tipping. The CR-Pro weighs 198
pounds and, even with attached casters, to avoid injury, it should not be
moved by less than two people. If lifted, such lifting should be
performed by professional movers or installers.