Installation guide

46
Series Number (0020,0010)
IS nnnnn
Image Number (0020,0013)
IS nnnnn
Samples per Pixel (0028,0002)
US 1
Photometric
Interpretation
(0028,0004)
CS MONOCHROME2
Rows (0028,0010)
US
Columns (0028,0011)
US
Bits Allocated (0028,0100)
US 8 or 16
Bits Stored (0028,0101)
US 8,16
High Bit (0028,0102)
US 7, 12, 15, or 16
Pixel
Representation
(0028,0103)
US 0
Pixel Data (7FE0,0010)
OW
3.3.4 TRANSFER Syntax Selection Policies
Only the DICOM Implicit Little Endian Transfer Syntax is supported.
3.4 Association Acceptance Policy
The Radlink CR Pro Image Transfer AE does not accept association requests.
4 Radlink CR Pro MWL AE Specification
The Radlink CR Pro MWL AE provides Standard Conformance to the following
DICOM V3.0 SOP Classes.
Table 6: Supported Meta SOP Classes
SOP Class SOP Class UID Role
Modality Worklist Query Find 1.2.840.10008.5.1.4.31
SCU
4.1 Association Establishment Policies
4.1.1 General
The Radlink CR Pro View Pro application issues a request to retrieve
Modality Worklist. It initiates an association to the Modality Worklist SCP
through Radlink CR Pro MWL AE component.
The maximum PDU size allowed is 16KB.