STEP 2: Calibration Procedure
Calibration of the CR PRO is required to compensate for any hardware
differences and should be performed at least once a month, depending on
usage. Recommended: A 14x17 inch cassette must be used for calibration. If a
smaller size cassette is used, banding will be apparent at the both sides of
scanned images when a larger cassette is used. Remove any grid that might be
present.
1. Expose Plate to Xray
Note: Expose the cassette on the side without the Radlink Label)
If distance is:
o 72” (measured from bottom of X-ray unit) at 70 Kvp, 30 MAs
o 40” (measured from bottom of X-ray unit) at 70 Kvp, 9 MAs
2. Rotate the cassette 180 degrees and expose with the same technique
used above.
3. Load Cassette Scan
Insert the exposed cassette in the CR Pro machine. (Figure 4)