Owner`s manual

36
FREQUENCY MEASUREMENTS
Since frequency and time are reciprocals of each other,
when you know one (frequency or time), you can easily
solve for the other (
f=
, then
t=
). Therefore, if you
measure the time period of a recurrent waveform as 0.2
ms, the frequency is 5 MHz (or ).
Note
: Since a DC voltage does not have a frequency,
you cannot make frequency measurements on a DC
voltage.
1. Set the input coupling switch to
GND
.
2. Connect the
GROUND
(0 volt) wire to the signal
source’s ground or reference level.
3. Rotate the
Zero Level
to vertically center the trace
on the grid overlay.
Note
: If the trace is widened by stray interference,
ground the probe body near the point being mea-
sured. The trace position is the voltage reference
line and all voltage measurements are read in
respect to this line. (Do not adjust the vertical
positioning control after the reference is estab-
lished.)
4. If it was necessary to ground the tip (see Step 3
note), remove the tip from the ground or reference
voltage. This readies the tip so you can use it to
make a measurement.
5. Set the input coupling switch to
AC
.
6. Click on the vertical cursor icon to
display the vertical cursors. The
Time
:
box appears below the
grid display.
1
t
1
f
1
0.2
ms
22-310.fm Page 36 Friday, August 6, 1999 12:26 PM