Product Specs

Volume adjustment
1. Press [VOL] button to adjust volume level.
2. When the volume class is , it mean that the alarm sound is off.
3. When power up, the sound volume level is the third class.
Calibration
1) Press [CH] button to select the channel for calibrate.
2) Press [CAL] button to enter the calibration mode, CAL flash, press [▼] or [▲] button to set reading error.
3) Press [CAL] button again to exit the calibration mode.
The errors adjust range is -4~4F.
Wireless control
1. Press [ ] button to turn on or off the wireless function. When power on, transmission function is on.
Receiver control
1. Press [ ] button to turn on the unit.
2. Press [ ] button to turn on the backlight 20 seconds.
3. Press [ ] button and hold for 3 seconds to turn off the unit.
4. Press [VOL] button to adjust volume level.
5. When the unit alarm, press any button to stop the alarm sound. This situation will send back to the transmitter and mute the alarm
sound.
Code rematch
1. The transmitter has already been synchronized with the receiver at ex-factory setting. If rematch is needed, press and hold [ ]
button for 5 seconds. Rematch mode would start and “Sync” icon will appear in the display.
2. Press and hold [VOL] button for 5 seconds to start rematch.
3. The units would restore the code automatically and return to normal mode.
NOTE:
1. Clean the probe immediately with cleaning wipes after each measurement to avoid cross contamination.
2. Do not expose thermometer to temperatures over 50C/ 122 F or use in ovens, otherwise incorrect readings or damage to the
thermometer may result.
ERROR SYMBOLS:
Symbol
Description
Action required
Full battery
Low battery
Replace the battery with the same type
Battery level indication
(75%/50%/25%)
EXT POWER
Power adapter connecting
CAL
Calibrating
SET TEMP
Fan connecting
LLL.L
The reading is out of low range(-58 F -50C)
Keep the measurement above low range
HHH.H
1.Sensor short circuit
2.The reading is out of high range (572 F/300C)
1.Return the thermometer for repair
2. Keep the measurement below high range
NO PROBE!
1.The sensor open circuit
2.The sensor probe is not connect
1.Return the thermometer for repair
2.Connect the sensor probe correctly
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment complies with the FCC RF radiation exposure limits set forth for an uncontrolled environment. This
equipment should be installed and operated with a minimum distance of 20 centimeters between the radiator and
your body. This device complies with Industry Canada Licence-exempt RSSs. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference,
including interference that m
ay cause undesired operation of the device. Le présent appareil est conforme aux CNR d’Industrie Canada
applicable aux appareils radio exempts de licence. L’exploitation est autorisée aux deux conditions suivantes : (1)
l’appareil ne doit pas produire de brouillage, et (2) l’utilisateur de l’appareil doit accepter tout brouillage
radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement.
Changes or Modifications not expressly approved by the party responsible could void the user’s authority to operate
this device.