User manual
1. DESCRIPTION AND OPERATION OF THE PRODUCT
1.1 Application of the Product
Radioactivity detector RADEX RD1008 evaluates the radiological situation according
to the value of the ambient dose equivalent H * (10) of ionizing photon (gamma - and X-ray)
radiation (hereinafter - the dose rate), the value of the flux density of beta- radiation (hereinafter
referred to as the flux density) and acumulates the ambient dose equivalent of gamma-
radiation (hereinafter - the dose).
1.2 Technical specifications
o Range of dose rate, µSv/h ….…………………………..……..……………..from 0,1 to 999
o Range of flux density measurement, 1/(cm
2
* min) …..……….……………. from 6 to 999
o Range of dose measurement ................................................. from 0,01 µSv to 350 mSv
o Range of energies of registered:
1) photon ionizing radiation, MeV ………………………….………….. from 0,05 to 3,0
2) beta radiation, MeV……………………………….……….…..………from 0,05 to 3,5
o Random error at a confidence probability of 0,95 (reproducibility of results)
1) dose rate, not more,% ……………………………………….…………...... ± (15+3/H)
where H - the dose rate in µSv/h.
2) flux density, not more,% ………………………………………………… ± (20+200/P)
where P - the flux density, 1/cm
2
* min.
3) the dose, not more,% ………………………..……………………………………..... 15
o Range of audio alarm thresholds
1) dose rate, µSv/h …….……………………………..…………………. from 0,2 to 1,20
2) flux density, 1/(cm
2
* min) ….……………………..…………………... from 10 to 120
o Time of measurement:
1) dose rate, s …………………………………………………..…………..…from 2 to 21
2) flux density, s..…………………….……………………….…….……………..…….. 21
o Power supply: battery «AA» type …………………….…………….…….…………….. 1 pc.
o Time of continuous work h, minimum ………….……..…………..….….………………..950
o Prouct dimensions height x width x thickness, mm, ……………………………140x64x26
o Weight (without batteries), kg, max ……………….……….……………….…….........0,175
¾ The measurement time depends on the dose rate. Increase of the number of
measurement cycles leads to the increase of reliability of indications.
¾ Time of continuous operation of the product specified for the battery with a capacity
of 3100mAh, with the natural background level not more than 0,30 µSv/h, not more than
5 1/(cm
2
* min), and factory settings of the product.
Factory settings:
Profile - «MEASURE»;
Dose rate threshold – «0,30 µSv/h»;
Flux density threshold – «10 1/cm
2
* min»;
Background –«OFF»;
Options: AUDIO – «QUIET»;
VIBRA-CALL– «OFF»;
DOSE – «OFF».
1.2.1 Customization tools:
¾ «MEASURE» or «SEARCH» modes
¾ setting of the alarm threshold, above which the alarm call or vibra signal turns on;
¾ calculation of the background values of dose rate of gamma-radiation and the flux
density of beta-particles;
¾ indication of background values on the display;
¾ indication on the display of excess of the measured values above background;
¾ accumulation of gamma-radiation dose during the observation period for two users;
¾ turning on / off the alarm;
¾ turning on/off audio;
¾ turning on/off vibra signal;
¾ turning on/off the display backlighting.
4