User manual

Digital B-RAD USER MANUAL
New World Technologies Inc. Page • 18
V2014.07.03
3.3.12 Tool Calibration
WARNING! Only qualified personnel with training in the safe operation of torque tooling and the Digital
B-RAD Tool System should operate this tool.
CAUTION! Do not calibrate at Target Torques that result in exceeding the Digital B-RAD Tool System’s
Torque Range. Severe tool damage will occur.
CAUTION! Calibration should only be done by a Qualified Calibration Technician. Improper use of the
calibration function will result in tool damage.
This function allows the operator to access the calibration values for the Digital B-RAD.
These values should only be modified by a Qualified Calibration Technician and using a
Calibration Stand.
To calibrate the Digital B-RAD:
1. Press the “Centre” button.
Result: The Main Menu screen will be displayed (Figure 3.3.12-1).
2. Use the “Up Arrow” button and the “Down Arrow” button to highlight the “Setup”
option.
3. Press the “Centre” button to select the “Setup” option.
Result: The Setup Menu screen will be displayed (Figure 3.3.12-2).
4. Use the “Up Arrow” button and the “Down Arrow” button to highlight the “Calib”
option.
5. Press the “Centre” button to select the “Calib” option.
Result: The Calibration screen will be displayed (Figure 3.3.12-3).
Figure 3.3.12-1: Main Menu Screen
Figure 3.3.12-2: Setup Menu Screen
Figure 3.3.12-3: Calibration Screen
6. Do one Torque Cycle.
7. Press the “Centre” button to highlight the Actual Torque (Figure 3.3.12-4).
Figure 3.3.12-4: Calibration Screen Actual Torque Highlighted