User guide

Quantity One User Guide
5-12
Fig. 5-10. Step Tablet dialog box.
The dialog box will indicate whether the displayed form is for the
transmissive or reflective step tablet, depending on whether you are in
transmissive or reflective mode.
Note: In transmissive scanning, the scanner is calibrated against a transmissive
step tablet. In reflective scanning, the scanner is calibrated against a
reflective step tablet, unless the reflective calibration strip is uncalibrated.
If the reflective calibration strip does not have step tablet values
associated with it, you can use the default values in the form.
In the Tablet Serial Number field, you can type in the serial number for the
step tablet values you are entering. (The reflective step tablet will probably
not have a serial number.)
Your calibration strip will have a clear plastic cover. If the quantity offset
value for the clear cover is included in your calibration strip package, you can
enter this number in the Quantity Offset field. Otherwise, use a value of 0.045.