User guide

Quantity One User Guide
7-12
application with the same filter and light settings, it will not auto
recalibrate.)
Calibration Report
To print out a calibration report each time the densitometer calibrates, click on
the Calibration Report checkbox.
7.6 Acquire the Image
Note: Before scanning in transmissive mode, make sure the white balance
region of the scanning area is not covered or obstructed in any way.
To begin to scan, click on the Acquire button. The scanned image will begin to
appear in the scanning window, line by line.
To interrupt a scan, click on the Stop button. A message will ask you to
confirm the interrupt, and then you will be asked if you want to keep the
partial scan. This feature is useful if you overestimated the size of the area
you selected.
After the scan is complete, a window will open displaying the scan image, at
which point you can analyze and save it.
Note: The image will open with a default file name that includes the date, time,
and (if applicable) user name. However, unless you have selected Auto
Save After Scan, the file will not be saved until you select Save or Save As
from the File menu.
7.7 Other Options
Oversample
This feature allows you to scan at the maximum resolution of the scanner and
then use spatial averaging to create an image with lower resolution. This can
result in better images at lower resolution—however, it takes longer to scan.