7-10
Fig. 7-9. Step Tablet dialog box.
In the step tablet dialog box, type the serial number for the printout into the
Tablet Serial Number field.
The Quantity Offset field does not apply in the GS-800. This value should be
set to zero.
Next, enter the values for the transmissive step tablet in the appropriate fields
under the Diffuse column. After the step tablet is scanned, the software will
associate each density value with its corresponding segment on the step
tablet.
The step tablet density values do not need to be entered each time you
calibrate. Once they have been entered and saved, they will be automatically
recalled when the calibration strip is scanned.
When you are finished filling out the Step Tablet Values form, click on OK.