All lines must be terminated by an ASCII line feed <LF>, EOI or
both.
Write data may be in either decimal, hexadecimal, or binary format.
Non-decimal numbers must be preceded with "H" for hex or "B" for
binary.
Definition of
Commands
PDATAOUT
command
Syntax: PD[ATAOUT] <address>[.<ports>][,<address>
[.<ports>],...]
<address>::= Module address of the 1260-14 (1-12)
<ports>::= One or more consecutive ports to return data from.
A single port is specified as a decimal number from 0 to 11. A
group of ports is specified as two decimal numbers from 0 to 11
separated by a hyphen, (-), with the lower numbered port to the left
and the higher numbered port to the right. For example, the
command string PD 1.3-5 would return values for ports 3, 4 and 5.
Description: The PDATAOUT command will cause the module
specified by the logical address to return the data associated with
each requested port. The format of the return data is as follows:
<address>. 1260-14 DIGITAL INPUT/OUTPUT MODULE
<CR><LF>
<address>. [<port>: <data>,...]<CR><LF>
<address>.END<CR><LF>
where:
<address>::= A three digit module address. (001-012)
<data>::= From 0 to 256 pieces of data in either
decimal, hexadecimal or binary format.
<port>::= A 2-digit number specifying the port to
associate with the data that follows.