Specifications
User Manual 3155
Specifications A-3
Ramp Time Range 10 µs to 1 s, 3 digits, ±0.1%
Sweep
Description Sample Clock sweeps continuously from start to stop, at a
rate defined by the sweep time. More complex sweep modes
and types can be generated using the FM mode in
conjunction with the FM composer program.
Type Linear or logarithmic
Direction Up or down, depending on the start and stop setting
Range 50 S/s to 100 MS/s
Time 1 ms to 1000 s, 7 digits, ±0.1%
Advance Automatic, triggered, gated or software command
Marker
Output and Level Same as SYNC output.
Position Programmable for selected frequency
OPERATING MODES
Normal Continuous waveform is generated
Triggered Each input cycle generates a single output cycle.
Gated External signal enables generator. First output cycle
synchronous with the active slope of the triggering signal.
Last cycle of output waveform always completed
External Burst Preset number of up to 128K cycles stimulated by an
internal, or external trigger. This mode is not available in
Sequence mode.
Trigger Sources
External
Input Front panel BNC
Level TTL
Slope Sensitivity Positive or negative, programmable
Frequency 5 MHz to DC
Internal
Range 100 mHz to 2 MHz
Resolution 7 digits
Accuracy 0.1%
Software SCPI command
Backplane TTLTrig0 through TTLTrig7; STAR
System Delay
Trigger to waveform output 1 Sample Clock+120 ns