Astronics Test Systems Operation 3-15
sets the frequency to 10.7 MHz
(10.7e6), amplitude to 2
V, offset to
1 V, and duty cycle to 30%.
outp 1 Turns on the output.
The above is an example of a full utilization of the A
pply command,
including the frequency, amplitude, offset,
and duty cycle
parameters for a standard square wave. You may use the A
pply
command in a similar manner for other standard or
arbitrary
waveforms.
You may use the Apply command on a partial set
of the available
parameters of a waveform.
To leave out a parameter, place no
characters between the corresponding comma separators. For
example, you may specify the frequency and duty cycle of a square
wave while leaving the amplitude and offset at the default values,
as follows:
*rst Restores factory defaults.
appl:squ 12.7e6,,,45 Selects the standard
square wave as
the active function,
and simultaneously
sets the frequency to 12.7 MHz, leave
s
the amplitude
at the default value,
leaves offset at the default value, and
sets the duty cycle to 45%.
outp 1 Turns on the output.
In a similar manner, you may use the Apply command
with other
standard waveforms to set some parameters while leaving others at
their default values.
Gen
erating
Standard
Waveforms
Each standard waveform is built into the 3152B in a
lookup table or
equation. Ten standard function shapes are available:
Sine
Triangle
Square
Pulse
Ramp
Sinc
Exponential Decaying Pulse
Gaussian Pulse
Noise
DC
Every time you select a standard function
, the 3152B retrieves the
data points from a lookup table or calculates
them from equations,
and then places them into waveform memory.
Use the following commands to select one of the standard