Astronics Test Systems Operation 3-9
events:
init:cont off
Turns Continuous mode off, changing to
an interrupted mode. If you
have not
selected a specific interrupted run mode
since power-up
, the generator will
automatically be placed in Triggered
mode.
trig:retr 1 Turns on Re-triggered mode.
trig:retr:del <value> Sets the re-trigger delay time. The re-
trigger delay is measured from the last
point of the waveform cycle to the first
point of the next waveform cycle.
<trigger event> Starts the re-trigger generator. Y
ou may
select the source of trigger events from a
number of inputs, such as
software trigger,
front-panel input, or VXIbus
backplane
trigger lines.
The waveform will continue to re-trigger unless you
change the run
mode or turn off
power. Use the following commands to restore
continuous run mode:
trig:retr 0 This turns off Re-triggered mode. T
he
generator will revert to the T
riggered run
mode.
init:cont on This
removes the 3152B from interrupted
run mode and reverts to C
ontinuous
mode.
Gated Run Mode
In Gated mode, the output remains at a specific DC level until
a
valid event opens the gate and waveform output begins.
At the end
of the last output cycle, the output assumes a DC level equal to the
amplitude of the last point of the waveform.
Only triggers from hardware sources can open or close the gate
.
Use VXIbus backplane trigger lines or the front-
panel trigger input
as the gating control.
You may set trigger detection for either a rising or falling edge
, or a
high or low level.
There are two selectable modes for opening the gate:
Transition (edge): A transition in the same direction as the
programmed slope toggles the gate open or closed.
Level:
The gate remains closed as long as the trigger signal is
at the selected logic state (high or low), based on the trigger
level setting. The gate
opens when the trigger signal exceeds
the trigger level setting.