3152B User Manual Publication No. 980935 Rev. M
3-8 Operation Astronics Test Systems
not previously select a specific interrupted
mode, the generator will automatically
be
placed in Triggered mode.
Continuous Run
Mode
Upon power-up, the 3152B defaults to Continuous mode
and
automatically starts generating waveforms which
appear at the
output connector as the output relay is turned on. Use init:cont on
to return to Continuous mode from any interrupted mode.
Triggered Run
Mode
In Triggered mode, the output remains at a specific DC level until
a
trigger event initiates a single output cycle
(see “Selecting the
Trigger Source” for details about trigger parameters).
Each time a transition occurs
at the trigger input, the 3152B
generates one complete output cycle. At the end of the
waveform
cycle, the output assumes
a DC level that is equal to the amplitude
of the first point of the waveform.
If you have not in
voked any interrupted modes since applying
power to the 3152B, you may use the following command to turn off
Continuous mode and default to Triggered mode:
init:cont off
Turns off Continuous mode, changing to
an interrupted mode. If you have
not
selected a specific interrupted run
mode
since power-up, the generator will
default
to Triggered mode.
If you have used other interrupted modes since power-
up, then
send the following command sequence to guarantee a change to
Triggered mode:
init:cont off Turns Cont
inuous mode off in case this
was the most recent run mode.
trig:retr off
Turns Retriggered mode off in case this
was the most recent interrupted run mode.
trig:gated off
Turns Gated mode off in case this was the
most recent interrupted run mode.
trig:burs off
Turns Burst mode off in case this was the
most recent interrupted run mode.
Re
-Triggered Run
Mode
Re-triggered mode is a special case of Triggered mode where,
after
a valid trigger event, the generator automatically self-
issues triggers
separated by a programmable delay. The delay i
nterval is
measured from the end point of the waveform to the start point
of
the next waveform cycle. Use the following commands
to place the
instrument in re-triggered mode and to program the delay between