3152B User Manual Publication No. 980935 Rev. M
3-6 Operation Astronics Test Systems
outp 1 Enable output relay to turn output on
volt <value> Set up the amplitude level
volt:offs <value> Set up the offset level
Set up your oscilloscope to
observe that the 3152B generates a
sine waveform with the following properties:
Frequency: 1 MHz
Offset: 0 V
Amplitude: 5 V
The output of the 3152B is calibrated for signals applied to a 50
load. If your amplitude is twice as high as expected, then
the 3152B
output may not be properly terminated. In this case, either
add a
50 termination to the cabling or change your oscilloscope settings
so that its input uses the built-in 50 input termination. Also,
note
the i
nteraction between amplitude and the offset as described
below.
Amplitude
-Offset
Interaction
Amplitude and offset may be programmed freely as long as the
following relationship is observed:
pVpOffset
Amplitude
+ 16
2
(-8V to +8V peaks)
Amplitude-offset combinations
outside the above limits will generate
"settings conflict" errors.