User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Installation 2-7
Table 2-1, Valid and Invalid IP Addresses for Subnet Mask 255.255.255.0
IP Address Comment
123.234.45.211 Valid.
123.234.45.213 Valid. The first three numbers match the previous IP address. The fourth number
must be a unique number in the range of 1 to 254.
123.202.45.214 Invalid. Second number does not match the previous IP addresses. The first
three numbers must match on all IP addresses with subnet mask 255.255.255.0.
123.234.45.0 Invalid. The first three numbers are valid but the fourth number cannot be 0.
123.234.45.255 Invalid. The first three numbers are valid but the fourth number cannot be 255.
TIP
To find out the network settings for your computer, perform
the following steps:
For Windows 98/Me/2000/XP
1. Open a DOS prompt.
2. Type IPCONFIG.
3. Press <Enter>.
If you need more information, you can run ipconfig with the
/all option by typing IPCONFIG /all at the DOS prompt. This
shows you all of the settings for the computer. Make sure you
use the settings for the LAN adapter you are using to
communicate with the LAN device.
For Windows 95
1. Open a DOS prompt.
2. Type WINIPCFG.
3. Press <Enter>.
Select the Ethernet adapters you are using to communicate
with the Ethernet device from the drop-down list.
Connecting to the
USB Port
The 3100M-3152B has a front panel USB connector that allows
connection to a
USB memory device. This USB port has only o
ne
purpose
, to allow the downloading of
waveform data directly from a
USB memory device
without going through
the VXI controller. The
waveform data has to be stored on the
USB memory device
in a
special format. The negotiation between the
USB memory device
and
the instrument is automatic
.
The application program can then select
the loaded memory segment for use in an application.