User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems 3152B Specifications
A-9
Pulse Generator Waveform Characteristics
Operation The 3152B has a special mode where the
instrument type is transformed to operate as a
digital pulse generator. When this mode is
selected, the operation of the arbitrary waveform
and its outputs are disabled and possibly, arbitrary
waveforms are overwritten
Programmability 1. All pulse parameters, except rise and fall times,
may be freely programmed within the selected
pulse period provided that the ratio between the
period and the smallest incremental unit does not
exceed the ratio of 1,000,000 to 1 (4,000,000 to 1
with the memory option), thus the specifications
below do not show the maximum limit as each
must be computed from the above relationship.
2. Rise and fall times may be freely programmed provided that the ratio between the rise/fall time
and the smallest incremental unit does not exceed
the ratio of 100,000 to 1.
3. The sum of all pulse parameters must not exceed the pulse period setting
Pulse State On or Off. On generates pulse output; Off
generates 0 Vdc
Pulse Mode Single or double, programmable
Polarity Normal, inverted, or complemented
Period 80 ns minimum, programmed with 4 ns increments
Delay 0 ns min; 1e6 s max (2e6 s max with option 1)
Pulse Width 4 ns minimum; 1e3 s max
Double Pulse Delay 0 ns minimum; 1e3 s max
Rise/Fall Times 0 ns minimum; 1e3 s max (actual min = <5 ns)
High Time 0 ns minimum
Amplitude Window 10 mVp-p to 16 Vp-p
Low Level -8 V to +7.992 V
High Level -7.992 V to +8 V
Half-Cycle Waveform Generator Characteristics
Description Half Cycle waveforms are the same as the
standard functions except waveforms are
generated as a sequence of half cycles. The delay
between the half cycles is programmable
Function Shape Sine, Triangle, Square
Frequency Range 0.01 Hz to 1 MHz
Phase Start Range 0° to 360.0° (Sine and triangle only)
Start Phase Resolution 0.1°
Duty Cycle Range 0% to 99.99% (Square only)
Duty Cycle Resolution 0.1%
Run Modes Continuous, Triggered
Delay Between Half Cycles 200 ns to 20 s (Applies to continuous run mode
only)
Resolution 20 ns