User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
A-6
3152B Specifications Astronics Test Systems
Sequenced Waveforms Generator Characteristics
Description Segments may be linked and repeated in a user-
selectable order. Segments are advanced using
either a command or a trigger
Sample Clock Range
Continuous Run Mode 100 mS/s to 250 MS/s (300 MS/s, typically at
25°C)
All Other Run Modes 100 mS/s to 225 MS/s (240 MS/s, typically at
25°C)
Resolution 11 digits
Accuracy and Stability Same as reference
Advance Modes
Automatic Sequence Advance No trigger required to step from one segment to
the next. Sequence is repeated continuously per a
pre-programmed sequence table.
Stepped Sequence Advance Current segment is sampled continuously until a
trigger advances the sequence to the next
programmed segment and sample clock rate.
Single Sequence Advance Current segment is sampled the specified number
of repetitions and then idles at the end of the
segment. Next trigger samples the next segment
the specified repeat count, and so on.
Mixed Sequence Advance Each step of a sequence can be programmed to
advance either a) automatically (Automatic
Sequence Advance), or b) with a trigger (Stepped
Sequence Advance)
Sequencer Steps 1 to 16 k
Segment Loops 1 to 1 Meg
Minimum Segment Duration 500 ns
Minimum Segment Size in a Sequence 10 points
Custom Sequence Creation Software ArbConnection software allows instrument control
and creation of custom sequences. Sequences are
built as tables and downloaded to the instrument
Download Modes Normal, using SCPI commands; Fast, using binary
data (<250 ms / 100 steps)
Modulated Waveform Generator Characteristics
General
Description Using this mode of operation, one may select from
a list of built-in modulation schemes.
Carrier Waveform (CW) Sinewave
Modulation Source Internal
Run Modes Off (outputs CW), Continuous, Triggered, Delayed
Trigger, Re-trigger, Burst and Gated
Interrupted Modulation Carrier Idle Mode On or Off, programmable
Run Mode Advance Source Front panel TRIG IN, TTLTRG(0-7), Internal
trigger, Software commands
Trigger Delay
(Trigger input to modulation output) [(0; <100 ns to 20 s) + system delay]
Resolution 20 ns
Error 6 sample clock cycles + 150 ns