3152B User Manual Publication No. 980935 Rev. M
7-48 Adjustments and Firmware Update Astronics Test Systems
CAUTION
Before clicking on the Update button in the next step
below, be ready to observe the Fail LED on the 3152B,
and follow the instructions carefully.
9. Click the Update button and then immediately watch t
he Fail
LED on the 3152B front panel. Wait for the LED to come on
before proceeding. This will typically take between 5 and 20
seconds. If the Fail LED does not come on within two minutes,
turn off power to the VXIbus chassis and contact the factory for
further instructions.
10. After the Fail LED comes on, wait until it turns off or until at least
one minute elapses.
11. By this time, a download success message will have appeared
(Figure 7-16). Click the OK button. Then click the Exit button in
the updater window.
Figure 7-17, Successful Firmware Update
12. Turn off power to the VXI chassis
. Wait at least two seconds,
and then turn it back on.
13. Run the Resource Manager (Resman).
14. Send a *IDN? command to the 3152B. The
response will be a
text string stating the model number and firmware version, as in
this example: