3152B User Manual Publication No. 980935 Rev. M
7-46 Adjustments and Firmware Update Astronics Test Systems
Updating Message
-
b
ased Firmware
Before you attempt to update the firmware of your message-based
instrument (Models 3152B and 3100M-3152B), first check the revision
level of the product. Each firmware update was done for a reason and
therefore, if you want to update the firmware
to address a specific
problem in your system, check the readme file that is associated with
the update to see if an update will solve your problem. There are a
number of ways to check the revision level of your firmware:
1. Use a SCPI command from an external utility. You can read the
firmware version by sending the following query:
SYST:INFO:FIRM:VERS?
The response is a string showing the firmware version, e.g.,
1.18.
2. Using ArbConnec
tion, select the General/Filters panel from the
System tab and click on the Firmware Version button. The
General/Filters panel with the firmware revision indication is
shown in Figure 7-3.
The computer that will be used for performing the update must have
the 3152B Plug&P
lay driver installed. If it does not, then you may
download the current version from the Astronics Test Systems web
site (search for 3152B plug&play).
Use the following procedure to upgrade the firmware o
nce you
determine that a firmware upgrade will
improve the performance of
your instrument.
1. Install the 3152B in a VXIbus
chassis and apply power to the
chassis.
2. Run the VXI resource manager (Resman.exe).
3.
Copy the new 3152B object file to a folder on the hard disk. The
object file has a fil
ename containing the model number and
version number. For example, the file "ri3100-v1.27.bin" is for
Model 3100 (covering Models 3100M, 3100R, 3152A, and
3152B) and is version 1.27.
4.
Launch "updater.exe". To locate it, navigate to the installation
path of the 3152B Plug&Play driver. The path is typically:
C:\Program Files\IVI Foundation\VISA\WinNT\ri3152Bb
In this folder is the "firmware updater" folder, containing the file
"updater.exe". Double-click "updater.exe" to launch the updater.
5. The updater window will appear as shown in Figure 7-13. In the