Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-43
Channel
1 – High-
Frequency
Response
Equipment: Oscilloscope, waveform generator, x10 attenuator
Preparation:
1. Configure the waveform generator as follows:
Frequency: 1MHz
Waveform: Square wave
Amplitude: 1.6V
2. Connect the waveform generator channel 1 output to
the 3202A, channel 1, input A.
3. Connect the 3202A channel 1 output through the x10
attenuator to the oscilloscope.
4. Set oscilloscope to DC coupling and 50 termination.
Adjustment:
1. Adjust RV1, RV3 and C10 for transition time between
18ns and 22ns and lowest aberration.
Channel 2
High-
Frequency
Response
(3202A Only)
Equipment: Oscilloscope, waveform generator, x10 attenuator
Preparation:
1. Configure the waveform generator as follows:
Frequency: 1MHz
Waveform: Square wave
Amplitude: 1.6V
2. Connect the waveform generator channel 1 output to
the 3202A, channel 2, input A.
3. Connect the 3202A channel 1 output through the x10
attenuator to the oscilloscope.
4. Set oscilloscope to DC coupling and 50 termination.
Adjustment:
1. Adjust RV4, RV6 and C27 for transition time between
18ns and 22ns and lowest aberration.