3152B User Manual Publication No. 980935 Rev. M
7-42 Adjustments and Firmware Update Astronics Test Systems
3. Click the “Store” button. Since the computer screen gives no
visual response to the store operation, you may wish to get a
visual indication by observing a brief flash of the front-
panel
“Access” indicator (Model
3152B) or “Link” indicator
(Model 3100) as you click the “Store” button.
4. Click the “Close” button.
5.
Exit the ArbConnection software by clicking the “X” in the upper
right-
hand corner of the main window. This completes the
calibration procedure.
Calibrating the
3201A/3202A
Calibration of the 320
1A/3202A consists of adjusting the flatness of
the square waveform shape for the best combination of high frequency
response and minimum overshoot/undershoot. These adjustments are
required only during initial factory calibration or after repairs have been
made to the related circuitry.
Channel 1
Low-
Frequency
Response
Equipment: Oscilloscope, waveform generator, x10 attenuator
Preparation:
1. Configure the waveform generator as follows:
Frequency: 1kHz
Waveform: Square wave
Amplitude: 1.6V
2. Connect the waveform generator channel 1 output to
the 3202A, channel 1, input A.
3. Connect the 3202A channel 1 output through the x10
attenuator to the oscilloscope.
4. Set oscilloscope to DC coupling and 50 termination.
Adjustment:
1. Adjust RV2 for best combination of flatness and rise
time of the square waveform shape.
Channel 2
Low-
Frequency
Response
(3202A Only)
Equipment: Oscilloscope, waveform generator, x10 attenuator
Preparation:
1. Configure the waveform generator as follows:
Frequency: 1kHz
Waveform: Square wave
Amplitude: 1.6V
2. Connect the waveform generator channel 1 output to
the 3202A, channel 2, input A.
3. Connect the 3202A channel 1 output through the x10
attenuator to the oscilloscope.
4. Set oscilloscope to DC coupling and 50 termination.
Adjustment:
1. Adjust RV5 for best combination of flatness and rise
time of the square waveform shape.