3152B User Manual Publication No. 980935 Rev. M
7-38 Adjustments and Firmware Update Astronics Test Systems
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 35 for a DMM reading of 17.678 mV
±0.5 mV
1 Meg Mode: Adjust CAL:SETUP 35 for a DMM reading of 35.355 mV
±
1 mV
Setup 52
3 V Amplitude, Post Attenuators - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 3 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 36 for a DMM reading of 1.06066 V
±10 mV
1 Meg Mode: Adjust CAL:SETUP 36 for a DMM reading of 2.12132 V
±
20 mV
Setup 53
500 mV Amplitude, Post Attenuators - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
2. Connect
the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 500 mV
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 37 for a DMM reading of 176.78 mV
±2 mV
1 Meg Mode: Adjust CAL:SETUP 37 for a DMM reading of 353.55 mV