User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
7-36 Adjustments and Firmware Update Astronics Test Systems
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 50 mV
Offset: 1 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 31 for a DMM reading of 17.678 mV
±2 mV
1 Meg Mode: Adjust CAL:SETUP 31 for a DMM reading of 35.355 mV
±4 mV
Setup 48
1 V Amplitude, Amplifier Out - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 1 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 32 for a DMM reading of 353.55 mV
±3 mV
1 Meg Mode: Adjust CAL:SETUP 32 for a DMM reading of 707.11 mV
±6 mV
Setup 49
500 mV Amplitude, Amplifier Out - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
2. Connect the 3152B output to the DMM inp
ut. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz