3152B User Manual Publication No. 980935 Rev. M
7-34 Adjustments and Firmware Update Astronics Test Systems
(for the 50 calibration; omit this during the 1 Meg portion), Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 5 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 27 for a DMM reading of 1.7678 V
±20 mV
1 Meg Mode: Adjust CAL:SETUP 27 for a DMM reading of 3.5355 V
±
40 mV
Setup 44
1 V Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 1 V
Offset: 1 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 28 for a DMM reading of 353.55 mV
±3 mV
1 Meg Mode: Adjust CAL:SETUP 28 for a DMM reading of 707.11 mV
±
6 mV
Setup 45
500 mV Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation: