Astronics Test Systems Adjustments and Firmware Update 7-33
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 50 mV
Adjustment:
50 Mode: Adjust CAL:SETUP 25 for a DMM reading of 17.678 mV
±0.5 mV
1 Meg Mode: Adjust CAL:SETUP 25 for a DMM reading of 35.355 mV
±
1 mV
Amplitude
Adjus
tments
(
Modulation)
The amplitude adjustments (modulation) ensure that
AC levels are
within the specified range. Use this procedure if you suspect that the
amplitude accuracy is an issue for the CW modes.
Setup 42
10 V Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 10 V
Mode: Modulation
Adjustment:
50 Mode: Adjust CAL:SETUP 26 for a DMM reading of 3.5355 V
±35 mV
1 Meg Mode: Adjust CAL:SETUP 26 for a DMM reading of 7.0711 V
±70 mV
Setup 43
5 V Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination