User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-31
±1 mV
Setup 37
50 mV Amplitude, Amplifier Out Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 50 mV
Offset: 250 mV
Adjustment:
50 Mode: Adjust CAL:SETUP 21 for a DMM reading of 17.678 mV
±0.5 mV
1 Meg Mode: Adjust CAL:SETUP 21 for a DMM reading of 35.355 mV
±
1 mV
Setup 38
3 V Amplitude, Post Attenuators - Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 3 V
Adjustment:
50 Mode: Adjust CAL:SETUP 22 for a DMM reading of 1.0607 V
±10 mV
1 Meg Mode: Adjust CAL:SETUP 22 for a DMM reading of 2.1213 V
±20 mV