Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-29
Frequency: 1 kHz
Output: On
Amplitude: 100 mV
Offset: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 16 for a DMM reading of 35.355 mV
±0.5 mV
1 Meg Mode: Adjust CAL:SETUP 16 for a DMM reading of 70.711 mV
±1 mV
Setup 33
50 mV Amplitude, Amplifier In Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 50 mV
Offset: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 17 for a DMM reading of 17.678 mV
±0.5 mV
1 Meg Mode: Adjust CAL:SETUP 17 for a DMM reading of 35.355 mV
±1 mV
Setup 34
1 V Amplitude, Amplifier Out - Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 1 V