3152B User Manual Publication No. 980935 Rev. M
7-28 Adjustments and Firmware Update Astronics Test Systems
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 1 V
Offset: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 14 for a DMM reading of 353.55 mV
±3 mV
1 Meg Mode: Adjust CAL:SETUP 14 for a DMM reading of 707.11 mV
±6 mV
Setup 31
500 mV Amplitude, Amplifier In Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 1 V
1.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
2. Configure the 3152B as follows:
Frequency: 1 kHz
Output: On
Amplitude: 500 mV
Offset: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 15 for a DMM reading of 176.78 mV
±2 mV
1 Meg Mode: Adjust CAL:SETUP 15 for a DMM reading of 353.55 mV
±4 mV
Setup 32
100 mV Amplitude, Amplifier In - Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range: 100 mV
2. Connect the 31
52B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows: