3152B User Manual Publication No. 980935 Rev. M
7-26 Adjustments and Firmware Update Astronics Test Systems
Offset +1 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 10 for a DMM reading of +1 V ±5 mV;
Note and record the DMM reading.
1 Meg Mode: CAL:SETUP 10 for a DMM reading of +2 V ±10 mV;
Note and record the DMM reading.
Setup 27
-1 V Offset, Output Amplifier Out
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Connect the 315
2B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV
Offset -1 V
Output: On
Adjustment:
1. 50 Mode: CAL:SETUP 11 for a DMM reading of -1 V ±5
mV; note reading.
2. 1 Meg Mode: CAL:SETUP 11 for a DMM reading of -2 V ±10
mV; note reading.
3. Repeat Setup 26 and Setup 27 until errors are balanced
between the steps.
Amplitude
Adjustments
The amplitude adjustments ens
ure that the AC levels are within the
specified range. Use this procedure if you suspect that amplitude
accuracy is an issue.
Starting with 3152B firmware revision 1.18, a 1 Meg calibration table is
available for calibration. This calibration omits the 50 feedthrough
termination and thus requires adjustment to twice the setup voltage as
for the normal 50 calibration. It is recommended that the 50
calibration is done first. Next, select 1 Meg calibration by using the
selection button in the upper right hand corner of the calibration panel
(see Figure 7-11).
Setup 28
10 V Amplitude, Amplifier In - Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion), Dual