User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-25
(for the 50 calibration; omit this during the 1 Meg portion), Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV
Offset -5 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 43 for a DMM reading of -5 V ± 25 mV
1 Meg Mode: CAL:SETUP 43 for a DMM reading of -10 V
±
50 mV
Setup 25
-7 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
4. Configure the DMM as follows:
Function: DCV
Range: 10 V
5.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
6. Configure the 3152B as follows:
Amplitude: 20 mV
Offset -7 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 42 for a DMM reading of -7 V ± 35 mV
1 Meg Mode: CAL:SETUP 42 for a DMM reading of -14 V
±
70 mV
Setup 26
+1 V Offset, Output Amplifier Out
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV