7-24 Adjustments and Firmware Update Astronics Test Systems
Amplitude: 20 mV
Offset +7 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 49 for a DMM reading of +7 V ± 35 mV
1 Meg Mode: CAL:SETUP 49 for a DMM reading of +14 V ± 70 mV
Setup 22
-1 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV
Offset -1 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 45 for a DMM reading of -1 V ± 5 mV
1 Meg Mode: CAL:SETUP 45 for a DMM reading of -2 V ± 10 mV
Setup 23
-3 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 10 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV
Offset -3 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 44 for a DMM reading of -3 V ± 15 mV
1 Meg Mode: CAL:SETUP 44 for a DMM reading of -6 V
±
30 mV
Setup 24
-5 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination