3152B User Manual Publication No. 980935 Rev. M
7-22 Adjustments and Firmware Update Astronics Test Systems
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Output: On
Amplitude: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 7 for a DMM reading of 0 V ± 5 mV
1 Meg Mode: Adjust CAL:SETUP 7 for a DMM reading of 0 V ± 10 mV
Offset Adjustments
The offset adjustments en
sure that the DC offsets are within the
specified range. Use this procedure if you suspect that offset accuracy
is an issue.
Starting with 3152B firmware revision 1.18, a 1 Meg calibration table is
available for calibration. This calibration omits the 50 feedthrough
termination and thus requires adjustment to twice the setup voltage as
for the normal 50 calibration. It is recommended that the 50
calibration is done first. Next, select 1 Meg calibration by using the
selection button in the upper right hand corner of the calibration panel
(see Figure 7-11).
Setup 18
+ 1 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Amplitude: 20 mV
Offset +1 V
Output: On
Adjustment:
50 Mode: CAL:SETUP 46 for a DMM reading of +1 V ± 5 mV
1 Meg Mode: CAL:SETUP 46 for a DMM reading of +2 V
±
10 mV
Setup 19
+3 V Offset Output, Amplifier In