Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-21
50 Mode: Adjust CAL:SETUP 4 for a DMM reading of 0 V ± 20 mV.
1 Meg Mode: Adjust CAL:SETUP 4 for a DMM reading of 0 V ± 40 mV.
Setup 15
Base Line Offset, Amplifier Out Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Mode: Modulation
Output: On
Amplitude: 1 V
Adjustment:
50 Mode: Adjust CAL:SETUP 5 for a DMM reading of 0 V ± 5 mV
1 Meg Mode: Adjust CAL:SETUP 5 for a DMM reading of 0 V ± 10 mV
Setup 16
Base Line Offset, Amplifier In Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Output: On
Amplitude: 6 V
Adjustment:
50 Mode: Adjust CAL:SETUP 6 for a DMM reading of 0 V ± 20 mV
1 Meg Mode: Adjust CAL:SETUP 6 for a DMM reading of 0 V
±
40 mV
Setup 17
Base Line Offset, Amplifier Out - Arbitrary
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter