3152B User Manual Publication No. 980935 Rev. M
7-20 Adjustments and Firmware Update Astronics Test Systems
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Function Standard
Waveform Sine
Amplitude 0 V
Frequency 1 MHz
Adjustment:
50 Mode: Adjust CAL:SETUP 2 for a DMM reading of 0 V ±20
mV
1 Meg Mode:
Adjust CAL:SETUP 2 for a DMM reading of 0 V
±40 mV
Setup 13
Pre-Amplifier Offset, Amplifier Out
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Function Standard
Waveform Sine
Amplitude 0 V
Frequency 1 MHz
Adjustment:
50
Mode: Adjust CAL:SETUP 3 for a DMM reading of 0 V ± 5 mV
1 Meg Mode: Adjust CAL:SETUP 3 for a DMM reading of 0 V ± 10 mV
Setup 14
Base Line Offset, Amplifier In Modulation
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion)
, Dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
termination (omit termination for 1 Meg calibration).
3. Configure the 3152B as follows:
Mode: Modulation
Output: On
Amplitude: 6 V
Adjustment: