Astronics Test Systems Adjustments and Firmware Update 7-19
Amplitude 2 V
Wave: Square
Adjustment:
1. Adjust CAL:SETUP 1 for a counter reading of 10 MHz ± 2 Hz
Base Line Offset
Adjustments
The base line offset adjustments assure that the AC signal is
symmetrical around the 0V line. Use this procedure if you suspect that
there is a base line accuracy issue.
Starting with 3152B firmware revision 1.18, a 1 Meg calibration table is
available for calibration. This calibration omits the 50 feedthrough
termination and thus requires adjustment to twice the setup voltage as
for the normal 50 calibration. It is recommended that the 50
calibration is done first. Next, select 1 Meg impedance mode by using
the selection button in the upper right hand corner of the calibration
panel (see Figure 7-11).
Figure 7-11, Selecting 1M Calibration Mode
Setup 12
Amplifier Offset, Amplifier In
Equipment: DMM, BNC to BNC cable, 50 feedthrough termination
(for the 50 calibration; omit this during the 1 Meg portion), dual
banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 100 mV
2.
Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50 feedthrough
Select 1M calibration mode