3152B User Manual Publication No. 980935 Rev. M
7-18 Adjustments and Firmware Update Astronics Test Systems
been minimized.
Oscillator
Adjustments
Use this procedure to adjust the reference oscillator
and gated
oscillator. These
oscillators determine the accuracy of the output
frequency and timing functions so,
if you suspect that there is an
accuracy issue, proceed with the calibration of the oscillators. Note
that (Setup 50 MHz) is marked as a factory adjustment and, therefore,
it is not normally required during nor
mal calibration cycles except
when the gated oscillator accuracy does not meet th
e published
specification limit or after a repair has been executed on this circuit.
(Setup 50MHz)
50 MHz Gated Oscillator Adjustment
Equipment: Counter, Function Generator, BNC to BNC cables
Preparation:
1. Configure the counter as follows:
Termination: 50 DC
Function: TI A -> B
Slope B: Negative
2. Connect the 3152B output to the oscilloscope input
3.
Connect an external function generator to the front panel
TRG/PLL connector
4. Using ArbConnection, prepare a
nd download the following
waveform:
Wavelength: 100 points
Waveform: Pulse: Delay = 0.01 %,
Rise/Fall Time = 0 %, High Time = 99.99 %
5. Configure the 3152B as follows:
Function Mode: Arbitrary
Run Mode: Triggered
Retrigger Mode: On
Retrigger Delay: 20 µs
6. Using an external function generator, manually trigger the
3152B
Adjustment:
1. Set CAL:SETUP 0
2. Adjust C10 for a period of 20 µs ±5%
(Setup TCXO)
10 MHz TCXO Frequency
Equipment: Counter, BNC to BNC cables
Preparation:
1. Configure the counter as follows:
Function: Freq A
Termination: 50
2. Connect the 3152B output to counter channel A.
3. Configure the 3152B as follows:
Frequency: 10 MHz
Ref Source: Internal
Output: On