Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-17
2. Set the oscilloscope channel impedance as follows:
Channel 1: 1 M
Channel 2: 50
3. Connect a BNC “T connector to oscilloscope channel 1.
4.
Connect one end of the “T” connector to the function generator
output.
5.
Connect the other end of the “T” connector to the TRG/PLL input
of the 3152B.
6. Connect the 3152B output to oscilloscope channel 2.
7. Configure the 3152B as follows:
Function Mode: Arbitrary
Wavelength: 1000 points
Waveform: Square
Run Mode: PLL
Adjustment:
1. Adjust CAL:SETUP 62 for counter reading of -250 ns to 250 ns.
(Setup 11)
Phase Fine “0”
Equipment: Counter, Function Generator, "T" Connector, BNC to BNC
cables
Preparation:
1. Prepare the external function generator waveform as
follows:
Amplitude: 5 Vp-p
Waveform: Square
Frequency: 10 kHz
2. Configure the counter as follows:
Channel 1 input impedance: High
Channel 2 input impedance: 50
Function: TI A -> B
3. Connect a BNC “T” connector to counter channel A
4. Connect one end of the “T” connector to the function
generator output.
5. Connect the other end of the “T” connector to the
TRG/PLL input of the 3152B, using a 50 terminator at
the 3152B connector.
6. Connect the 3152B output to counter input B.
7. Configure the 3152B as follows:
Function Mode: Arbitrary
Wavelength: 1000 points
Waveform: Square
Run Mode: PLL
Fine Phase Offset: +30°
Adjustment:
1. Adjust CAL:SETUP 61 for a counter reading of 8.28 µs
to 8.38 µs.
2. Repeat Setup 10 and Setup 11 until the errors have