7-16 Adjustments and Firmware Update Astronics Test Systems
(Setup 8)
290 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 290 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 57 for a DMM reading of 0 V, ±100 mV
PLL Adjustments
The PLL operation is basic to the sample clock generator. Correct
operation of the PLL circuit en
sures the accuracy of the frequency
path as well as the accuracy of the PLL phase. Use this procedure if
you suspect that there is a frequency and/or PLL accuracy issue.
(Setup 9)
PLL Preparation
Equipment: No equipment required for this step.
Preparation:
1. Configure the 3152B as follows:
Function Mode: Arbitrary
Wavelength: 1000 points
Waveform: Square
Run Mode: PLL
2. Set CAL:SETUP 61
3. Set CAL:VAL 120
4. Set CAL:SETUP 63
Adjustment: None required
. This step is part of the preparation for
Setup 10 below.
(Setup 10)
Phase “0”
Equipment: Digital Oscilloscope, Function Generator, BNC "T"
Connector (one male to two females), BNC to BNC coaxial cable
Preparation:
1. Prepare the external function generator waveform as follows:
Amplitude: 5 Vp-p
Waveform: Square
Frequency: 10 kHz