Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-15
SCLK: 220 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 53 for a DMM reading of 0 V, ±100 mV
(Setup 5)
240 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 240 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 54 for a DMM reading of 0 V, ±100 mV
(Setup 6)
260 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 260 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 55 for a DMM reading of 0 V, ±100 mV
(Setup 7)
277 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 277 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 56 for a DMM reading of 0 V, ±100 mV