3152B User Manual Publication No. 980935 Rev. M
7-14 Adjustments and Firmware Update Astronics Test Systems
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 160 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 50 for a DMM reading of 0 V, ±100 mV
(Setup 2)
180 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 180 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 51 for a DMM reading of 0 V, ±100 mV
(Setup 3)
200 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows:
SCLK: 200 MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 52 for a DMM reading of 0 V, ±100 mV
(Setup 4)
220 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function: DCV
Range: 1 V
2. Configure the 3152B as follows: