Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-13
NOTE
Parameters that are adjusted only during Factory
Calibration are enclosed in parentheses. All other
parameters are calibrated during Factory Calibration or
Periodic Calibration.
In the panel, each parameter to be calibrated is labeled with a number
from 1 to 57 except the 50MHz and TCXO adjustments in the PLL
group. Adjustments should be carried out exactly in sequential order.
The numbers that are associated with each adjustment are identified
by a setup number in
the title of each of the adjustments in the
following procedure.
Adjustment values have a range of 0 through 255, with a center
setting of 128. If you are not able to calibrate within the specified
range, there is either a problem with the way you are measuring the
parameter or a problem with the instrument. In either case, do not
leave any adjustment in its extreme setting but center the adjustment
and resolve the problem before proceeding with the calibration.
Note that in the following procedures,
although configuration of the
3152B is done automatically, configuration details
are shown for
reference. There is no requirement to change the 3152B configuration
during the adjustment procedure except where specifically noted.
Note
After making calibration adjustments, it is essential
that you lock the calibration factors as described in
the section “Locking the Calibration Factors” at the
end of this procedure. This stores the calibration
factors in flash memory so that the instrument will
retain its calibration after power is turned off.
VCO Adjustments
The VCO controls the accuracy and linearity of the sample clock
generator. Correct operation of the VCO circuit ensures the accuracy
of the frequency path. Use this procedure if you suspect that there is a
frequency linearity and/or accuracy issue.
(Setup 1)
160 MHz SCLK
Equipment: DMM
Preparation: