3152B User Manual Publication No. 980935 Rev. M
7-12 Adjustments and Firmware Update Astronics Test Systems
Factory
Calibration
Warm
-up
Install the 3152B in a VXIbus chassis. Turn on chassis power and
allow the 3152B to warm up and stabilize for at least 30 minutes
before calibrating. If the instrument has been subjected to
temperatures outside the range of 25°C ± 5°
C , allow at least one
additional hour for the instrument to stabilize before calibrating.
Procedure
Initialization
1. Invoke ArbConnection.
2. Click on the Calibration tab on the Panels bar.
The Calibration
Panel shown in Figure 7-10 will appear.
Figure 7-10, Calibration Panel