3152B User Manual Publication No. 980935 Rev. M
1-10 Introduction Astronics Test Systems
VXI controller but take control via the LAN port to completely
bypass the backplane interface and control instrument functions
and parameters from a web page using LXI capabilities
. The
USB port is used for memory stick I/O
where waveform data
can be loaded directly to arbitrary waveform
memory. This
feature was specifically designed for security reasons where
breach of secret waveform data can endanger national security
if it falls to the wrong hands
. This way, the data is erased
immediately as soon as the instrument is turned off but can be
restored by a person holding a memory stick
with data. This
version can be expanded to a 3100M-3152B-
3152B for dual
channel configuration where two instruments ca
n be used
independently of each other or synchronized for master-
slave
operation. The 3100M carrier can accept
other instruments as
well although these are not described here.
3100M-3152B-3201A This version
includes the 3152B
waveform generator combined
with the 3201A Signal Amplifier
module. By connecting the output of the 3152B to the input of
the 3201A, higher drive capability is available
at the output of
the 3201A.
3100M-3152B-3202A This version
includes the 3152B
waveform generator combined with the dual-
channel 3202A
Signal Amplifier module.
The description in this manual ignores specific configurations and
provides details of the 3152B only. The
configuration chosen can
have a significant impact on the way that one programs the
instrument. Register based commands are
parsed and executed
using an external dll (3100R) but message-based commands are
parsed and executed by the internal CPU (3100M, 3151B, 3152B)
.
The supplied drivers provide access to
all functions for both
versions.
Specifications
Appendix A lists the instrument specifications. Specifications apply
under the following conditions:
Output terminated into 50 except where otherwise specified
30-minute warm up period
Ambient temperature range of 20°C to 30°C.
For temperatur
es outside the above temperature range,
specifications degrade by 0.1 % per °C.