3152B User Manual Publication No. 980935 Rev. M
7-8 Adjustments and Firmware Update Astronics Test Systems
Figure 7-7, Selecting 1M Calibration Mode
15. Disconnect the 50 termination load and connect the 3152B
output directly to the DMM input.
16. Set the DMM function to DC Volts.
17.
In the same manner as described above, adjust each calibration
item given in Table 7-3.Disconnect the 50
termination load and
connect the 3152B output directly to the DMM input.
Table 7-3, Calibration Values, Unloaded (>1MΩ)
Item in
Cal Panel
DMM
Function
Low Limit
(Volts)
Target
(Volts)
High Limit
(Volts)
12
VDC
-0.040
0.00000
0.040
13
VDC
-0.010
0.00000
0.010
14
VDC
-0.040
0.00000
0.040
15
VDC
-0.010
0.00000
0.010
16
VDC
-0.040
0.00000
0.040
17
VDC
-0.010
0.00000
0.010
18
VDC
1.990
2.00000
2.010
19
VAC
5.970
6.00000
6.030
20
VAC
9.950
10.00000
10.050
21
VAC
13.930
14.00000
14.070
22
VAC
-2.010
-2.00000
-1.990
23
VAC
-6.030
-6.00000
-5.970
24
VAC
-10.050
-10.00000
-9.950
25
VAC
-14.070
-14.00000
-13.930
26
VAC
1.990
2.00000
2.010
Select 1M calibration mode