7-6 Adjustments and Firmware Update Astronics Test Systems
Figure 7-6, Making an Adjustment
11. Use the up-arrow and down-arrow keys on the computer
keyboard to adjust the number up or down as required, while
observing the reading on the DMM
. Adjust the number until the
DMM reading is as close
as possible to the target value given in
Table 7-2.
12. When finished, verify that the DMM reading is
within the low and
high limits given in Table 7-2
. If you are not able to calibrate
within the specified range, there is either a problem with the way
you are measuring the parameter or a problem with the 3152B
.
In either case, do not leave any adjustment in its extreme setting
but center the adjustment to 128
and contact your nearest
service center for clarification and support.
13. Repeat steps 7 through 11 above for each row in Table 7-2.
Table 7-2, Calibration Values, 50 Load
Item in
Cal Panel
DMM
Function
Low Limit
(Volts)
Target
(Volts)
High Limit
(Volts)
12
VDC
-0.020
0.00000
0.020
13
VDC
-0.005
0.00000
0.005
14
VDC
-0.020
0.00000
0.020
15
VDC
-0.005
0.00000
0.005
16
VDC
-0.020
0.00000
0.020
17
VDC
-0.005
0.00000
0.005
18
VDC
0.995
1.00000
1.005
19
VDC
2.985
3.00000
3.015
20
VDC
4.975
5.00000
5.025
21
VDC
6.965
7.00000
7.035
22
VDC
-1.005
-1.00000
-0.995
Click the calibration adjustment number
Then position vertical cursor here