Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-5
Figure 7-4, Selecting the Calibration Function
Figure 7-5, Calibration Panel
In the Calibration Panel, each calibration item has a selection button
numbered from 1 to 57, except the 50MHz and TCXO adjustments.
For Periodic Calibration, only items 12 through 55 should be adjusted.
For each row in Table 7-2, do the following.
8. Verify that the DMM is set to the function indicated in the table.
9. In the Calibration Panel, select the item indicated in Table 7-2.
10. Click the calibration adjustment number (Figure 7-6)
. Using the
right arrow key on the computer keyboard, position the vertical
cursor immediately to the left of the right-most digit as shown.
Each item to be calibrated
has a numbered selection
button
Select “System”
Then select “Calibration”