Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-3
Warm
-up
Install the 3152B in a VXIbus chassis. Turn on the chassis power and
allow the 3152B to warm up and stabilize fo
r at least 30 minutes
before calibrating
. If the instrument has been subjected to
temperatures outside the range of 25°C ± 5°
C , allow at least one
additional hour for the instrument to stabilize before calibrating.
Procedure
1. Using a BNC coaxial cable and BNC-to-banana-plug adapte
r,
connect the 3152B front-panel output to the DMM input.
2. Set the DMM function to DC Volts.
3. Start the ArbConnection software. The
Startup & Communication
Options window will appear as shown in Figure 7-1.
Figure 7-1, Startup & Communication Options Window
4. Use the mouse to select “Detect Automatically”, “VXI MB”, and
“Communicate Only” as shown in Figure 7-1.
Then click the
“Communicate” button.
5. The ArbConnection main window will then appear (Figure 7-2).
Select these options
Then click the Communicate button